Environmental Science & Engineering - www.esemag.com - March 2005
Comments? send them to the editor.
Measuring dissolved oxygen in biological reactors
Almost all dissolved oxygen
analysers are calibrated using
air as the reference. At 25°C,
the saturation value of oxygen
in water is 8.4 ppm. However, oxygen
concentrations in a biological reactor
(aerobic and anaerobic zones) normally
range from 2.0 ppm down to zero.
Very few manufacturers recommend
checking the zero point, even though
the measurement range is much closer
to zero than the air calibration point.
There are four basic sensor technologies
available today, and their low range
measurement capabilities are as follows:
- Galvanic (membrane). The
anode and cathode are immersed in an
electrolyte, into which oxygen permeates
through a membrane. The galvanic
sensor converts oxygen into a voltage
(via a sacrificial anode) that is proportional
to the amount of dissolved oxygen.
Therefore, the sensor has an
absolute zero. Zero D.O. = Zero output.
- Galvanic (no membrane). The
electrodes of this sensor are exposed to
the wastewater, which is used as the
electrolyte. The sensor is calibrated in
oxygen saturated water. As water has
different pH and conductivity values to
that of wastewater, there can be no certainty
that the measurement, especially
below 2 ppm, is accurate.
- Polarographic (membrane).
This is sometimes called amperometric.
The anode and cathode are
immersed in an electrolyte, into which
oxygen permeates through a membrane.
It differs dramatically from a
galvanic sensor in that the anode has to
be polarized, after which a current
flows in the sensor. So at 0 ppm D.O.,
the sensor has an output which is offset
by the parent analyser. As the sensor
ages, its zero offset changes.
- Optical (Fluorescent). This is
sometimes called Luminescent. A certain
wavelength light is focused onto a
fluorescing material. which produces
light at a different wavelength. The
presence of oxygen slows or reduces
the amount of fluorescent light produced.
As light is constantly produced
in a bandwidth rather than an absolute
wavelength, there is no absolute zero.
However, as the measurement is frequency
(rate of change) based, there is
no drift as long as the signal strength is
reasonable. However, the detector sees
any light (sunlight, area lighting, etc.)
as a change in oxygen concentration, so
it must be calibrated and operated in an
almost black box environment. It must
be submersed into a reactor, sometimes
cannot normally be used in effluent
channels, and in some cases the sensor
is irreparably damaged by sunlight.
Development History
The Clark Cell principle for the
measurement of dissolved oxygen,
both galvanic and polarographic was
developed in the 1960s. Over the years
major advancements were made, such
as the use of pure electrodes rather
than plated ones, and electrochemically
cleaned membranes.
The optical measurement technique
has also been available for several
decades. But there is a major problem in
that the fluorophore is soluble in water.
So the material needs to be bonded to
another material, which impedes or
stops the degradation of the fluorescing
material. Only the Ruthenium (fluorophore)/
Silicon (bond) matrix does
this, and this sensor is patented and
licensed to one manufacturer.
Third Party Verification
While there have been numerous
tests and evaluations carried out in the
wastewater industry since the 1980s,
the only independent, scientific evaluation
of all sensor technologies was
performed in 2002/2003.
The Instrument Testing Authority
carried out testing of 10 D.O. analysers
in Decatur, Illinois over a three month
period. The comprehensively documented
report makes no recommendations or
draws conclusions, it simply reports on
the performance of the technologies.
The report can be found at
www.instrument.org
Operation in Biological Reactors
There are many myths associated
with accurate, reliable on line D.O.
measurement. So much so that there
are some who flatly refuse to use anything
but timers for control. Others
have tried ORP to control D.O., and
some have used nutrient monitors.
Some have simply given up and set
blowers to fixed settings.
A primary requirement in these
reactors is biological nutrient removal.
A common methodology is to control
D.O. at below 0.5 ppm so that nitrification
and denitrification occur simultaneously.
This requires D.O. analysers
to reliably and accurately measure
under 0.5 ppm.
Modern reactors are commonly
designed with two zones, aerobic and
anaerobic. This somewhat reduces the
reliance on D.O. analysis, as nitrification
and denitrification are bound to
occur. However, without D.O. control,
the process is inefficient, and the delay
in using nutrient results for control is
similarly inefficient. Using timers or
fixing airflows is even less efficient,
and results in increased sludge removal
and/or chemical costs.
Conclusion
A reliable dissolved oxygen analysis
system costs approximately 0.1%
of a biological reactor project cost, so
there is no valid reason for not purchasing
D.O. instrumentation for the
reactor, and certainly not the cheapest
available instrumentation. It is well
publicized that approximately 60% of
a plant’s operational cost is due to
blower electricity consumption, so a
reliable D.O. monitoring/control system
is quickly paid for in electrical
savings alone.
The selection of such a D.O. system
needs to be made based on accuracy,
long-term reliability and maintainability.
There are systems available that
are proven to operate accurately and
reliably for 12 months or more with no
maintenance whatsoever. Selecting the
right dissolved oxygen analyzer is
paramount in improving treatment and
lowering costs.
For further information, please email sales@spdsales.com.
See our home page on how to order your subscription. We regret we can
only accept orders from Canada.